2.
If using a sterilization tray, follow any additional instructions provided with the tray. Use
only trays that are compatible with EO.
3.
Double wrap the device (or tray) prior to sterilization.
4.
Sterilize the device using the parameters below.
P P r r e e c c o o n n d d i i t t i i o o n n i i n n g g
Temperature
Chamber Humidity
Vacuum Set Points
Time
E E x x p p o o s s u u r r e e
Concentration
Temperature
Time
Chamber Humidity
A A e e r r a a t t i i o o n n
Aeration Time
Temperature
5.
Allow the cable to cool to room temperature before reuse.
EN- 14
55 °C (131 °F)
70% RH
1.3 psia
30 minutes
725 mg/L, 100% EO
55 ± 2 °C (131 ± 3 °F)
1 hour
70% RH (50–80%) ± 5%
12 hours
55 ± 2 °C (131 ± 3 °F)