W W a a r r n n i i n n g g s s a a n n d d C C a a u u t t i i o o n n s s
1.
Read these instructions thoroughly before using the device.
2.
Before using this device, read the light source user manual for warnings, indications,
intended use, and other information about using the light source system.
3.
Federal law (United States of America) restricts this device to use by, or on order of, a
physician.
4.
Carefully unpack the device and check if any damage occurred during shipment.
5.
This device is shipped non-sterile. Clean and sterilize the device prior to the first use
and after every subsequent use. Follow the cleaning, disinfection, and sterilization
instructions provided in this user manual.
6.
Test the device function prior to use. If there is any sign of malfunction, the device
should not be used and returned to Stryker for repair evaluation.
7.
Do not abuse, pull, stretch, kink, puncture, or otherwise alter the cable. Doing so will
cause irreversible damage to the glass optical fibers, which will impair light transmission
through the cable.
8.
The surface temperature near the scope adapter and at the tip of the scope can exceed
41 °C if the light source is operated at high levels of brightness for extended periods of
time. The heated scope and adapter can cause burns to the patient, user, or
combustible materials.
9.
When operating the light source, never look into the following apertures or direct the
light emitted from the apertures toward another person:
•
the light cable connection on the light source (if the cable is not attached)
•
the end of the light cable (or scope adapter)
•
the scope tip
10. Never use this equipment in the presence of flammable or explosive gases.
11. Replace the cable if the light transmission is compromised due to damaged fibers or
connectors.
12. No modification of this equipment is allowed.
The warranty is void if any of these warnings or cautions are disregarded.