Catalog Numbers
RH4701-001, RH4701-002, RH4701-003, RH4701-004,
RH4701-005, RH4701-006, RH4701-007, RH4701-008,
RH4701-009
Indications For Use
Laryngeal mirrors are intended for use in surgical procedures to
examine the larynx.
How Supplied
V. Mueller devices are packaged as non-sterile. Cleaning and
sterilization must occur prior to use.
Limitations on Reprocessing
Repeated reprocessing has minimal effect on these devices.
End of life is normally determined by wear and damage due to
use. Repeated soaking in cleaning and disinfection solutions
may cause discoloration of the mirror.
Warnings
Devices shall be used in accordance with these instructions for
use. Read all sections of this insert prior to use. Improper use
of this device may cause serious injury. In addition, improper
care and maintenance of the device may render the device non-
sterile prior to patient use and cause a serious injury to the
patient or health care provider.
Cautions
If there are any variations between these instructions and either
your facility's policies and/or your cleaning/sterilization
equipment manufacturer's instructions, those variations should
be brought to the attention of the appropriate responsible
hospital personnel for resolution before proceeding with
cleaning and sterilizing your devices.
Use of device for a task other than that for which it is intended
will usually result in a damaged or broken device.
Prior to use, inspect device to ensure proper function, and
condition. Do not use devices if they do not satisfactorily
perform their intended function or have physical damage.
Only the cleaning and sterilization processes which are defined
within these instructions for use have been validated.
Use only neutral pH (6-8) detergent solutions.
Pre-processing Instructions
Initiate cleaning of device within 2 hours of use.
Transport devices via the institution's established transport
procedure.
Remove excess gross soil as soon as possible after use by
rinsing or wiping the device.
All devices must be processed in the completely open position
(i.e. flushports, jaws, etc.) to allow solution contact of all surfaces.
Manual Cleaning
1. Ensure all pre-processing instructions are followed prior to
cleaning.
2. Prepare the enzymatic / neutral pH detergent solution,
utilizing tap water with a temperature range of 27°C to 44°C
(81°F to 111°F), per manufacturer's instructions.
3. Place device in the open/relaxed position, with flush port
open, and completely immerse in the detergent solution and
allow device to soak for a minimum of 5 minutes. Actuate
all movable parts during the initiation of the soak time.
4. Using a soft bristled brush, remove all visible soil from the
device. Actuate device while brushing, paying particular
attention to hinges, crevices and other difficult to clean
en