Troubleshooting

5.2 Alarm Codes

5.2
Alarm Codes
Alarm Codes and Descriptions
The following alarm codes appear on the main display of the flow meter.
Letter Codes
S SPACE
R HI/LO
F FAULT
T TURB
M MEMRY
K MAKUP
I
Z
The displays shown below indicate where the Alarm Codes appear on the screen. Press
<UP> or <DOWN> Arrows to change screen views.
42
Alarm Code
Spacing
Sensor spacing may need adjustment
Rate
Flow above High setting or below Low setting
Fault
Three continuous seconds without new data update
Turbulence
Current turbulence percentage exceeds the alarm set point
Memory
Last valid reading for a selected interval during Fault condition
Makeup
In-Process Makeup occurred
The following alarm codes appear in the Datalogger status messages:
Vs Interface
Gas Vs exceeds Vs Interface set point
ZeroMatic
ZeroMatic signal fault
Alarm Codes
Description
FUT1010 IP65 NEMA 4X & IP66 NEMA 7 Gas Quick Start
Operating Instructions, 1/2013, A5E02639183-AC