Alignment; Assemble The Complete Limb; Adjust The Pylon Length - Blatchford LINX22L1S Manual Del Técnico Ortoprotésico

Ocultar thumbs Ver también para LINX22L1S:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 146

8 Alignment

The device will need to be adjusted for
the height of the user, and the footwear
to be used.
Follow these instructions to adjust the
length of the pylon in the device.
WARNING
Take care not to damage the
internal connecting cable between
the knee and ankle modules when
assembling or disassembling the
device.
8.1
Assemble the Complete
Limb
Ensure the connecting cable is not trapped,
carefully slide the knee onto the top of the
pylon tube with a slight twisting action (there
is no need to connect the cable at this stage).
Ensure that the pylon is fully inserted into the
knee housing then align the knee and ankle
and tighten the clamp bolt to 10 Nm Torque.
Take the measurement required for the user.
8.2

Adjust the Pylon Length

Undo the clamp screw and taking care not
to damage the connecting cable remove the
Knee module.
To release the pylon undo the 4 pyramid set
screws on the ankle module, carefully ease the
shin cable and grommet from the pyramid and
remove the pylon.
Knee
module
Connecting
cable
Extension
cable
Grommet
seal
Ankle
module
16
Clamp Screw
5
10 Nm
Pylon
Connecting
cable
Set
screws
938353V2APK1/1-0321