Maintenance and Troubleshooting
Display
Action
Blood glucose is below the defined hypoglycaemic (low blood glucose) level. See
Chapter 2, Unusual Blood Glucose Results.
A general marker was assigned to this test result.
A pre-meal marker was assigned to this test result.
A post-meal marker was assigned to this test result.
A pre-meal marker was assigned to this test result and the post-meal test reminder
has been activated.
The test strip may be damaged, not properly inserted, or was previously used.
Remove and reinsert the test strip or replace it if damaged or previously used.
This error could appear if the cap on the test strip container was not closed tightly.
The test strips may have been damaged due to improper storage or handling.
A meter or test strip error has occurred.
Repeat the blood glucose test. If a second E-3 code appears, perform a control test
with the control solution and a new test strip.
• If the control result is within the acceptable range, review the proper testing
procedure and repeat the blood glucose test with a new test strip.
• If the control result is not within the acceptable range, see the Understanding Out-of-
Range Control Results section in the chapter Control Tests.
In rare cases, an E-3 error code may indicate that your blood glucose is extremely
high and above the system's measuring range. If you do not feel like your blood
glucose is extremely high, repeat the blood glucose test. See the Unusual Blood
Glucose Results section in the chapter Blood Glucose Tests. If the E-3 code still
appears for your blood glucose test, contact your healthcare professional
immediately.
Not enough blood or control solution was drawn into the test strip for measurement or
was applied after the test had started. Discard the test strip and repeat the blood
Document Name: ART-07022590007
glucose or control test.
Document Number: 0000109119
Version: 1.0
This copy will not be updated. Verify the validity of this version in the appropriate electronic repository.
34
78120_07022590007_EN.indb 34
Lifecycle Status: Effective
4/23/2019 1:56:44 PM
Confidential
Valid from: 02-Aug-2019
Page 36 of (180) page(s)

Publicidad

loading