Balloon Inflation Volume
Do not exceed maximum inflation volume. Adhere to balloon inflation
parameters as shown in Figs. 1 and 2. Over-inflation of balloon may result in:
• Damage to vessel wall and/or vessel rupture.
• Rupture of balloon.
Maximum Inflation Volumes
CODA-2-10.0-35-120-46
CODA-2-10.0-35-140-46
CODA-2-9.0-35-100-32
CODA-2-9.0-35-120-32
50
45
40
35
30
25
20
15
10
5
0
0
5
50
45
40
35
30
25
20
15
10
5
0
0
5
INSTRUCTIONS FOR USE
NOTE: "Balloon" lumen is for inflation and deflation of balloon.
NOTE: "Distal" lumen is for wire guide insertion and withdrawal.
Catheter Size
Balloon Catheter Inflation Parameters
20
10
15
Inflation Volume (cc)
● = CODA-2-9.0-35-100[120]-32
Balloon Catheter Inflation Parameters
20
25
10
15
Inflation Volume (cc)
u = CODA-2-10.0-35-120[140]-46
DISTAL
Coda LP
25
30
35
40 45
Coda
30
35
40 45
3
Max. Volume
60 cc
60 cc
30 cc
30 cc
50
55
60 65
Fig. 1
50
55
60 65
Fig. 2