4 Performing a functional test
4.1 Checking the rotating unit mechanics within a test cycleConverting Instructions "ORTHOPHOS XG 3D
12
4
Performing a functional test
Performing a functional test
4.1
Checking the rotating unit mechanics within a test
cycle
Checking the rotating unit mechanics within a test cycle
The test cycle is executed without radiation. The test cycle is used to
check that the unit is functioning correctly and to ensure that a complete,
uninterrupted cycle is possible. In the event of increased resistance, the
rotating unit stops automatically.
1. Touch the PAN symbol at the top of the touchscreen.
XGPlus/XG 3D test cycle
The program group PAN is selected.
2. Press the T key on the control panel.
The program enters test cycle mode.
3. Press the release button.
The test cycle is started.
4. Check the mechanics of the rotating unit during the test cycle:
Can the cycle be executed without interruption?
Are there any unusual noises during the cycle?
5. If the cycle is interrupted or unusual noises can be heard, the
mechanics of the rotating unit are defective.
6. Inform the operator of any deficiencies.
NOTICE! The conversion may only be performed on a unit that is
intact.
7. Press the T key on the control panel again.
The program exits test cycle mode.
4.2
Checking the height adjustment mechanics
Checking the height adjustment mechanics
1. Use the Up/Down keys on the control panel to move the unit up and
down through its entire adjustment range.
2. Check the height adjustment mechanics:
Does the height adjustment function correctly?
Are there any unusual noises during the adjustment?
3. If the height adjustment does not function correctly or unusual noises
can be heard during the adjustment (e.g. a hammering noise which
changes with the speed), the mechanics of the height adjustment or
the height adjustment motor are defective.
4. Inform the operator of any deficiencies.
NOTICE! The conversion may only be performed on a unit that is
intact.