7 Updating the unit software and configuring the unit
Saving data to board DX88
40
Converting Instructions "ORTHOPHOS XG 3D
1. Call service routine S009.8 (saving/restoring DX88 data) and execute
the routine (see ORTHOPHOS XG 3D Service Manual, section
"Service → S009: Test step 8").
2. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
The unit software has now been updated and the unit is configured.
Following a restart, only error message E1 11 20 (unit is not
calibrated) should appear.
Sirona Dental Systems GmbH
ready
to ORTHOPHOS XG 3D" ORTHOPHOS XG 3D
D3352.092.08.05.02 079.2012
ready
63 14 541 D3352