6 Mechanical conversion
6.5 Removing the old X-ray tube assembly Converting Instructions "ORTHOPHOS XG 3D
22
3. Loosen the two rear screws (C) on the tube assembly.
4. CAUTION! The tube assembly is heavy!
Hold the tube assembly firmly in place, loosen the two front screws
(D) (3-4 turns) and remove the tube assembly toward the front.
IMPORTANT
The old X-ray tube assembly must not be reused.
Within the scope of the upgrade, it must be packed into the packaging
of the new X-ray tube assembly and returned to Sirona with the
enclosed return note (see chapter entitled Returning the X-ray tube
assembly to Sirona [ → 48]).