5.
perform a second spin.
6.
At the end of the spin, change the position of the calibration weight on the shaft as-
sembly bell as indicated in figure 27;
7.
Perform a third spin. This last calibration phase includes the execution of three
consecutive spins in automatic mode.
If the calibration has been successful, a permission acoustic signal goes off at the end of
the spin; otherwise, the E 2 message is temporarily displayed.
Notes:
- Once the procedure is finished, remove the calibration weight;
- Press the
- This calibration is valid for any kind of wheel.
Operator's Manual S 65 E - S 65 E C
button to interrupt the calibration procedure at any time.
26