Cleaning and Disinfection
The Modified Tri-Star Swivel Adapter should be thoroughly cleaned after each use. Scrub with a soft brush and use pH-
neutral detergent. Clean thoroughly to prevent traces of blood or debris from interfering with movement. Wipe assembly
with an antiseptic solution.
Do not steam sterilize!
Plastic components may be damaged by heat.
Manual Wash
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause damage to the devices.
• Channels and crevices found on this device require particular attention during cleaning.
• Pay special attention to the water quality used throughout reprocessing. Hard water can damage the surface of the
equipment. Avoid using hard water. Instead use purified water unless otherwise specified.
Limitations on reprocessing
• Repeated processing has minimal effects on these devices. Product life is normally determined by wear and damage due
to use.
• It is important to have Integra NeuroSpecialists perform routine inspections (twice yearly is recommended). See contact
information below.
INSTRUCTIONS
Containment/Transportation
• Follow health care facility protocol for safe containment and transport to the decontamination environment.
• It is recommended that devices are cleaned immediately after use.
Preparation for Cleaning
• Disassembly is required.
Disassembling Equipment
Completely remove the swivel adapter from any
components. Unlock the Tri-Star bar and remove
torque screw and Tri-Star bar.
Cleaning – Manual Equipment: Water, Neutral pH Detergent, Soft Bristle Brush, Towels
Method
1. Prepare neutral pH enzymatic detergent solution (e.g. Endozime® AW Triple Plus with APA (Ruhof), 1:128 ratio) according
to detergent manufacturer's instructions using lukewarm tap water.
2. Prepare equipment for soaking by disassembling removable parts and loosening connections.
3. Rinse equipment in warm water before placing into bath.
4. Completely soak equipment in water/detergent solution for 30 minutes maximum.
5. Clean thoroughly with a soft nylon bristle brush. NOTE – If possible, use a disposable brush.
6. Rinse in warm purified water until all visible substances and residual detergent are removed. NOTE – Make sure to give
special attention to hard-to-reach areas.
7. Thoroughly dry equipment with soft clean towels and use medically compressed air if needed, to dry channels, crevices
and lumens.
8. Inspect the equipment to make sure there is no visible organic debris or residue from cleaning agent.
Repeat process if any soil is detected.
Drying
• Products should be dry at this point. If wetness or excess liquid is detected, dry with a soft clean towel.
• Medically compressed air can be used if needed.
Inspecting for Cleanliness
Inspect the teeth of the starbursts, the area around the
locking screw, and the base of the swivel adaptor for
organic debris.