2. Apply the adaptor to the clamp and carefully inspect the starburst connection to see that it is properly meshed and
secure. See Illustration 3.
WARNING:
If the swivel adapter knob becomes tight before the starburst is properly meshed, the knob is incorrectly adjusted
for this skull clamp. Use a 5/64" hex wrench to turn the adjustment screw counter-clockwise four full turns to
reset the screw. See Illustration 1 for adjustment screw location. If the starburst fails to mesh correctly after the screw is
reset, the Tri-Star Modified Swivel Adapter may be incompatible with this particular clamp. See Section 2 for details.
3. With the adapter securely applied to the skull clamp, use a 5/64" hex wrench to turn the adjustment screw clockwise
until it is tight. See Illustration 4. Check the ratchet extension for movement.
Correct
Incorrect
Use 5/64" hex wrench
to perform adjustment.
The starbursts shown
in the connection here
have completely meshed
to positively prevent
movement of the skull
clamp base.
The starbursts shown in
this connection have failed
to completely mesh. Even
though the torque screw
is tight, movement at the
joint is detectible. See
note below for directions
to evaluate the condition.
Illustration 3
Illustration 4