Dräger SilverLine Instrucciones De Uso página 26

Tabla de contenido

Publicidad

EnglishEnglish
Service
Safety information
WARNING
Risk due to inappropriately reprocessed prod-
ucts
The product may be contaminated with infec-
tious agents.
Before service is performed and before the
product is sent back for repair, reprocess the
product in accordance with the chapter "Repro-
cessing".
WARNING
Risk if service is not performed regularly
Wear and material fatigue of the components
may lead to device failure and malfunctions.
Perform service at the specified intervals.
WARNING
Risk if service is not performed properly
Personal injury and property damage may oc-
cur if service is not performed properly.
Service must be performed by those user
groups that are assigned to the particular mea-
sure.
WARNING
Risk if maintenance is not performed properly
If the device is connected to the gas supply
during maintenance, there is a risk of personal
injury and property damage.
Before performing maintenance, disconnect all
gas connections from the gas supply.
26
Definition of maintenance concepts
Concept
Definition
Service
All measures (inspection, main-
tenance, repair) intended to
maintain or restore the function-
al integrity of a product
Inspection
Measures intended to determine
and assess the current state of a
product
Maintenance
Regular specified measures in-
tended to maintain the functional
integrity of a product
Repair
Measures intended to restore
the functional integrity of a prod-
uct after a failure
Inspection
Checks
Interval
Personnel respon-
sible
Inspection
Every
Service personnel
2 years
Maintenance
Compo-
Interval
Measure
nent
Sealing of
Every
Replace-
pressure
2 years
ment
reducer/
gas cylin-
der
O-ring for
Every
Replace-
NIST
4 years
ment
hose
Compo-
Interval
nent
Pressure
Every
reducer
6 years
Repair
Dräger recommends that all repairs are performed
by DrägerService and that only authentic Dräger re-
pair parts are used.
Disposal
This product must be disposed of in accordance with
national regulations.
Technical data
Ambient conditions
During operation
Personnel
Temperature
responsi-
Ambient pressure
ble
Service per-
Relative humidity
sonnel
During storage and transport
Temperature
Service per-
Ambient pressure
sonnel
Measure
Personnel
responsi-
ble
Major over-
Specialized
haul, re-
service per-
placement
sonnel
of mem-
branes and
sealing ele-
ments
5 to 40 °C
700 to 1100 hPa
Altitude up to 4600 m
5 to 95 %, non-condens-
ing
–20 to 60 °C
500 to 1100 hPa
Altitude up to 5520 m
Instructions for use SilverLine

Publicidad

Tabla de contenido
loading

Tabla de contenido