Dräger SilverLine Instrucciones De Uso página 27

Tabla de contenido

Publicidad

Relative humidity
5 to 95 %, non-condens-
ing
Pressure reducer
All specified values for volume, flow, and leakage
are expressed in STPD conditions.
Type
Single-stage fixed pres-
sure reducer
Permissible input pres-
sure range (p3 to p1)
Air, O
11 to 200 kPa x 100
2
*)
N
O, CO
11 to 60 kPa x 100
2
2
Nominal output pres-
4.5 kPa x 100, fixed set-
sure (p2)
ting
Air, O
2
p3 = 11 kPa x 100
4,5 +/- 0,5 kPa x 100
Q1 = 80 L/min
p1 = 200 kPa x 100
4,5 +/- 0,5 kPa x 100
Q1 = 80 L/min
2 *
N
O, CO
2
p3 = 11 kPa x 100
4,5 +/- 0,5 kPa x 100
Q1 = 65 L/min
p1 = 60 kPa x 100
4,5 +/- 0,5 kPa x 100
Q1 = 65 L/min
Nominal flow Q1
Air, O
80 L/min
2
*
N
O, CO
65 L/min
2
2
Performance data
See average flow char-
acteristics on pages 28
to 30.
Relief valve
Opening pressure
7.5 + 0.5 kPa x 100
Flow
At least 220 L/min Air
Instructions for use SilverLine
Inlet port
According to one of the
following standards:
DIN 477
BS 341
DIN EN 850 / CGA V-1 /
ISO 407 (pin index)
CGA V-1
NF E 29-650
NBN 226
GB 15383 (only O
ISO 5145
Outlet port
Optionally screw con-
nection NIST according
to ISO 18082
or
hose with screw connec-
tion NIST according to
ISO 18082
Weight
1.2 to 2 kg (2.7 to 4.4 lb)
(depending on the type
of connector)
High-pressure sensor
Measuring range
0 to 250 kPa x 100
Display accuracy
<0.5 % at 23 °C
Degree of protection
IP67
Pressure gauge
According to EN 837-1
Display accuracy
Class 2.5
Measuring range for
0 to 315 kPa x 100
Air, O
2
Measuring range for
0 to 150 kPa x 100
*
N
O, CO
2
2
Classification Medi-
cal Device Europe
UMDNS Code
Universal Medical De-
vice Nomenclature Sys-
tem – Nomenclature for
Medical Devices
*
CO
Only for USA
2
)
2
English
Class II b
13-323
27

Publicidad

Tabla de contenido
loading

Tabla de contenido