Dräger SilverLine Instrucciones De Uso página 43

Tabla de contenido

Publicidad

During storage and transport
Temperature
–20 to 60 °C
(–4 to 140 °F)
Ambient pressure
500 to 1100 hPa
(7.3 to 16 psi)
Altitude up to 5520 m
(18110 ft)
Relative humidity
5 to 95 %, non-condens-
ing
Pressure reducer
All specified values for volume, flow, and leakage are
expressed in STPD conditions.
Type
Single-stage fixed pres-
sure reducer
Permissible input pres-
sure range (p3 to p1)
Air, O
11 to 200 kPa x 100
2
(160 to 2900 psi)
N
O, CO
11 to 60 kPa x 100
2
2
(160 to 870 psi)
Nominal output pressure
4.5 kPa x 100 (72.5 psi),
(p2)
fixed setting
Air, O
2
p3 = 11 kPa x 100
4.5 +/- 0.5 kPa x 100
(160 psi)
(65.3 ±7.2 psi)
Q1 = 80 L/min
p1 = 200 kPa x 100
4.5 +/- 0.5 kPa x 100
(2900 psi)
(65.3 ±7.2 psi)
Q1 = 80 L/min
N
O, CO
2
2
p3 = 11 kPa x 100
4.5 +/- 0.5 kPa x 100
(160 psi)
(65.3 ±7.2 psi)
Q1 = 65 L/min
Instructions for use SilverLine
p1 = 60 kPa x 100
4.5 +/- 0.5 kPa x 100
(870 psi)
(65.3 psi ±7.2 psi)
Q1 = 65 L/min
Nominal flow Q1
Air, O
80 L/min
2
N
O, CO
65 L/min
2
2
Performance data
See average flow char-
acteristics on pages 44 to
46.
Relief valve
Opening pressure
7.5 + 0.5 kPa x 100
(101.5 + 7.2 psi)
Flow
At least 220 L/min Air
Inlet port
CGA V-1
Outlet port
Optionally screw connec-
tion NIST according to
ISO 18082
or
hose with screw connec-
tion NIST according to
ISO 18082
Weight
1.2 to 2 kg (2.7 to 4.4 lb)
(depending on the type of
connector)
High-pressure sensor
Measuring range
0 to 250 kPa x 100 (0 to
3626 psi)
Display accuracy
<0.5 % at 23 °C (73.4 °F)
Degree of protection
IP67
Pressure gauge
According to EN 837-1
Display accuracy
Class 2.5
Measuring range for
0 to 315 kPa x 100
Air, O
(0 to 4600 psi)
2
Measuring range for
0 to 150 kPa x 100
N
O and CO
(0 to 2200 psi)
2
2
Classification Medical
Class II b
Device Europe
UMDNS Code
13-323
Universal Medical De-
vice Nomenclature Sys-
tem – Nomenclature for
Medical Devices
English US
43

Publicidad

Tabla de contenido
loading

Tabla de contenido