2.2 Area of application
Our components perform optimally when paired with appropriate compon
ents based upon weight and mobility grades identifiable by our MOBIS clas
sification information and which have appropriate modular connectors.
The product is recommended for mobility grade 1 (indoor walk
er) and mobility grade 2 (restricted outdoor walker).
m°
kg
•
The maximum approved body weight is specified in the technical data
(see Page 19).
2.3 Environmental conditions
Allowable environmental conditions
Temperature range: –10 °C to +45 °C
Moisture: relative humidity: 20% to 90%, non-condensing
Unacceptable environmental conditions
Chemicals/liquids: fresh water, salt water, perspiration, urine, acids, soapsuds,
chlorine water
Solids: dust, sand, highly hygroscopic particles (e. g. talcum)
Other environmental conditions
UV light
Hydrolysis
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative humidity 20 % to
90 %, no mechanical vibrations or impacts
2.4 Service life
This product has been tested by the manufacturer with 2 million load cycles.
Depending on the patient's activity level, this corresponds to a service life of
2 to 3 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
Not resistant
Not resistant