► In case of changes in or loss of functionality, do not continue using the
product (see "Signs of changes in or loss of functionality during use" in
this section).
► Take any necessary measures (e.g. repair, replacement, inspection by
the manufacturer's customer service, etc.).
Signs of changes in or loss of functionality during use
Decreased forefoot resistance or changes in roll-over behaviour are notice
able indications of loss of functionality.
4 Scope of delivery
Quantity
1
1
Spare parts/options (not included in the scope of delivery)
Connection plate
Foot adapter with screw connection
Ankle block (exoskeletal design)
Screw connection for ankle block (exoskeletal design)
Ottobock special adhesive with hardener
5 Preparation for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
NOTICE
Grinding the prosthetic foot
Premature wear resulting from damage to the prosthetic foot
► Do not grind the prosthetic foot.
5.1 Installing the foot adapter or ankle block
The installation must be performed if the prosthetic foot is not equipped with
a foot adapter or ankle block.
Designation
Only for 1D10 without adapter and for 1D11
Designation
Instructions for use
Prosthetic foot
Reference number
2R8*, 2R31*, 2R54*
2Z22=M8x90, 2Z22=M10
636W28 or 636W18 and
2R14
2K34=*
636W19