Dräger Tropic plus Instrucciones De Uso página 21

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
Wall-mounted version on rail
 Attach the bracket and the pole (600 mm long) to a standard
rail (25x10 mm) on the wall.
 Fit Tropic plus onto the pole using the holder at the back of the
nebulizer and secure it in place with the locking screw.
 Slide the holding-down device (to secure the lid of the nebu-
lizer chamber) and the hinged arm onto the pole and position
them as required.
Trolley-mounted version
A
 Slide the retaining ring (A) onto the pole and screw tight.
Instructions for Use Tropic plus
 Fit Tropic plus onto the pole of the trolley using the holder at
the back of the nebulizer and secure it in place with the locking
screw.
 Slide the holding-down device (to secure the lid of the nebu-
lizer chamber), the hinged arm, and the bottle holder (for stan-
dard refilling systems) onto the pole and position them as
required.
Transport with trolley
WARNING
Risk of tipping over
For intrahospital transport push Tropic plus down to the
pre-mounted stop (maximum 70 cm) above the floor) and
move it to the preferred position.
Guide the device during transport.
CAUTION
Risk of crushing during assembly, start-up, and use of the
hinged arm.
English
21

Publicidad

Tabla de contenido
loading

Tabla de contenido