Replacing the quartz oscillator
 Screw the new quartz oscillator (C) into the nebulizer chamber
and check it for tight fit. The edge of the nebulizer chamber
must be flush with the quartz oscillator.

Replacing the fuse

Set the toggle switch at the back to "0" (off).
Carefully unlock the fuse insert on the device socket using a
screwdriver and remove it. Replace the fuse and put the fuse
insert back in by pressing it until it audibly clicks into place.
Only use the following type of fuse:
2 x IEC 60127-2/III for T 1 L 250 V; 5 x 20 mm
Fault – Cause – Remedy
Fault
Cause
Red light is lit.
Nebulizer chamber
not fitted correctly.
Nebulizer chamber
empty.
Leak in nebulizer
chamber.
Faulty quartz oscil-
lator.
Electronic fault.
Not enough nebuli-
Nebulizer fluid does
zation when set at
not match scale.
maximum output.
Resistance of filter
too high.
Condensate in
hose.
26
max.
min.
max. 134 °C
C
Remedy
Insert nebulizer
chamber correctly.
Be careful that it is
properly positioned.
Refill nebulizer
fluid.
Screw quartz oscil-
lator firmly into the
nebulizer chamber.
Replace quartz
oscillator.
Call DrägerService.
Refill nebulizer fluid
or remove some.
Replace filter.
Empty condensate.
Re-route hose in
such a way that
condensate can
flow back into the
nebulizer chamber.
Fault
Cause
Leak in hoses or in
the lid of the nebu-
lizer chamber.
No nebulization.
Mains switch to "0"
(off).
Mains plug not con-
nected.
Faulty mains cable. Replace mains
Mains fuse blown.
No nebulizer fluid in
nebulizer chamber.
Faulty quartz oscil-
lator.
Electronic fault.
Heated hose does
Plug of hose heater
not heat up.
not connected.
Faulty hose.
Electronic fault.
Reprocessing
Disassembly
 Disconnect the mains plug and the device plug.
 Disconnect the plug of the hose heater.
 Disconnect the hoses and the bacterial filter.
 Push the down-holding device up and remove the lid of the
nebulizer chamber. Detach the nebulizer chamber and empty
out the fluid. Observe the hospital hygiene regulations.
 Dispose of disposable articles (disposable hoses, medication
container and lid, containers of sterile water systems) in an
environmentally-friendly manner. Observe the hygiene regula-
tions.
 Do not disinfect or sterilize the filter.
 Replace and dispose of the filter as directed by the manufac-
turer.
Remedy
Fit lid tightly and
secure with clamp.
Connect hoses
firmly to nozzles.
Turn mains switch
to "I" (on).
Connect mains
plug.
cable.
Replace mains
fuse, use 2 x IEC
60127-2/III for T 1 L
250 V type.
Check voltage.
Check preselected
voltage
230/127/110 V
Refill nebulizer
fluid.
Replace quartz
oscillator.
Call DrägerService.
Connect plug.
Replace hose.
Call DrägerService.