Adjustment possibility of
the prosthetic foot
Shifting the prosthetic foot
in the test adapter in the
proximal direction
Shifting the prosthetic foot
in the test adapter in the
distal direction
1) In the 647H545 Quick Reference Guide, note the section "Dynamic
Optimisation" (see page 15).
2) CAUTION! See the instructions for use of the prosthetic compon
ents for the mounting torque values.
Use the distal and proximal set screws on the socket adapter for posteri
or or anterior placement of the prosthetic foot.
3) To shift the prosthetic foot in the test adapter, see the section "Shifting
the Prosthetic Foot" (see Page 23).
5.5 Finishing the Prosthesis
Required tools:
>
Torque wrench
1) Tighten all screws of the prosthetic components to the specified installa
tion torque values.
2) Verify proper functioning of the prosthesis.
6 Use
NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
► Check the product for damage prior to each use.
► Do not use the product if its functionality has been impaired.
► Take any necessary measures (e.g. repair, replacement, inspection by
the manufacturer's customer service, etc.).
24
Effect
Reducing springiness
upwards
Increasing springiness
upwards
Application
When the patient has the
feeling of falling forward or
wants a less sporty gait.
When springiness upwards
is excessive.
When springiness upwards
is not sufficient.