5 Preparation for use

CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
5.1 Installing the foot adapter
► Install the foot adapter on the prosthetic foot according to its instructions
for use.
5.2 Alignment
NOTICE
Grinding the prosthetic foot
Premature wear resulting from damage to the prosthetic foot
► Do not grind the prosthetic foot.

5.2.1 Bench Alignment

TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring apparatus,
743A80 50:50 gauge, alignment apparatus (e.g. 743L200 L.A.S.A.R. Assembly or
743A200 PROS.A. Assembly)
Perform the assembly and alignment of the prosthetic components in the alignment
tool according to the following specifications:
Sagittal plane
Heel height: effective heel height (shoe heel height - sole thickness of fore­
   
foot) + 5 mm
Exterior foot rotation: approx. 5°
   
14
Designation
Prosthetic foot
Bench alignment process
Reference number
Reference number
2R31=M8, 2R54=M8
2R14
2K34=*
2Z22=M8x90
636W28 or 636W18 and 636W19