CLASS II equipment
Insulation
class
Type B
Applied Part complying with the specified
Applied
requirements for protection against electrical
Part
shock according to IEC60601-1.
Sound
45 to 50 dB (A)
level
Degree of
The control unit, external power supply,
protection
motors and hand control are protected
according to IPX4.
IPX4 – The electrical system is protected
against water splashed against the
components from any direction.
The bed has no isolator (main switch). If the bed
needs to be electrically disconnected, it has to be
unplugged from the mains socket.
1556305-H
9.5 Environmental conditions
Storage and
Transportation
Temperature
-10°C to +50°C
Relative humidity
20% to 90% at 30°C - not condensing
Atmospheric
pressure
Allow the product to reach operation temperature
before usage:
•
Warming up from minimum storage temperature
might take up to 24 hours.
•
Cooling down from maximum storage
temperature might take up to 24 hours.
Technical Data
Operation
+5°C to +40°C
800 hPa to 1060 hPa
25