Bed With 24 V External Power Supply; Installing Side Rails; Verso Ii Side Rail - Invacare Medley Ergo Manual Del Usuario

Ocultar thumbs Ver también para Medley Ergo:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 57
Invacare® Medley® Ergo
8.
IMPORTANT!
The hand control, control unit and motors
are protected according to IPX4.
– A locking cam must be used on the control
box to guarantee the IP protection.
Install the locking cam over the plugs in the control box.
9.
Attach the mains cable in the strain relief that is
attached to the bed frame.
10. Before taking the bed into use, check that all parts run
smoothly and that no motor cables are hanging on the
floor or being squeezed while moving bed sections.

3.6.1 Bed with 24 V external power supply

1. Fasten the short cable A to the control box (a "click"
must be heard)
2. Attach the cable (as described in step 7 in the Wiring
section).
3. Connect the short cable B to the transformer cable C.
4. Connect the transformer 24 V D to the 230 V main
socket.

3.7 Installing side rails

WARNING!
Risk of personal damage
There is a risk of entrapment or squeezing, while
assembling or disassembling the side rail.
– Pay attention to your fingers.
– Follow instructions carefully.
– After each assembly, check that all fittings are
properly tightened and that the side rail has
the correct function.
12
Removable 3/4 length side rails
WARNING!
Risk of entrapment
There's a risk of entrapment or suffocation
between mattress support, side rail and bed end.
– When using removable side rails, always make
sure that the distance between the bed end
and the side rail's handle upper edge is less
than 6 cm in the head end and more than 32
cm in the foot end of the bed.

3.7.1 Verso II side rail

Mounting
The mounting of the side rail must be done with the
side rail in locked position.
1. Place both brackets of the side rail on the frame of the
bed, with the locking button A facing towards the foot
end of the bed.
2. Ensure that the bracket near the head end of the bed is
placed encompassing the lifting pole support as shown
in detail B.
3. For beds equipped with bed ends with Sanne style
fitting, it is beforehand necessary to temporarily
disengage the locking pin C from the head bed end
to place the bracket.
a. Pull the locking pin C and rotate 1/4 turn to lock it
in the disengaged position.
b. Position the bracket under the locking pin as
described in step 2.
c. Rotate the locking pin and allow it to engage with
the bed end.
4. After the brackets are positioned correctly on the frame
of the bed, fix the side rail by firmly tightening the
finger screws D.
Dismounting
1. Untighten the finger screws and lift the side rail of the
bed frame.