Technical Data - Endress+Hauser RIA250 Manual De Instrucciones

Ocultar thumbs Ver también para RIA250:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 194

9. Technical data

General details
Unit function
Application
Process
display, power
supply, alarm
contactor
Operating and
Measurement
system
principle
construction
Measurement
system
Input
Input Typees
Measurement
range
Lineariation
Integration time 1 s
Process display for panel mounting
The display receives an analogue signal and shows the
corresponding value on the display. The analogue output
transmits the displayed value as either a current of
voltage. Two presettable limit values monitor the
measured value for any infingement of the preset
conditions and control the two output relays. Transmitters
connected can be directly powered by the unit.
The analogue signal connected is digitalised, analysed
and indicated in the display. A digital/analogue convertor
makes a proportional current or voltage signal available
for additional peripheral equpment at the output terminals.
Microcontroller controlled display with LED display,
analogue input, analogue output, limit alarm relays and
loop power supply.
Voltage, current, resistive thermometer (RTD),
thermocouple (TC)
Voltage:
±100 mV; max. voltage
±10 V;
R
: 1 MOhm
i
Current:
0/4 to 20 mA;
R
: 5 Ohm
i
RTD: Pt100: -200° to +850 °C (DIN EN60751)
Ni100: -60° to +180 °C (DIN 43760)
Sensor current: approx. 250 µA, pulsed
Connection: 2-, 3-, 4-wire
Cable compensation: 40 Ohm
TC: Type T: -270 to +400 °C Type B:
Type J: -210 to +1200 °C Type N: -270 to +1300 °C
Type K: -200 to +1372 °C Type U: -200 to +600 °C
Type R: -50 to +1800 °C Type L: -200 to +900 °C
Type S:
0 to +1800 °C Type W3:
Type T, J, K, R, S, B, N to DIN EN60584;
Type U, L to DIN 43710;
Type W3, W5 to ASTME988-96
Possible using max. 32 points
Process display
±5 V
without damage ±50 V
Over-range +200 mA
0 to +1820 °C
0 to +2315 °C
Type W5:
0 to +2315 °C
73

Publicidad

Tabla de contenido
loading

Tabla de contenido