Interrupting the Exposure
If the exposure key is prematurely released, the expo-
sure is interrupted and terminated.
The kV/mA value and the LED over key R blink at the
Multitimer.
The exposure time passed until the interruption is
shown.
Press the R key on the Multitimer twice.
Check patient positioning and repeat the exposure.
Automatic Exposure Blockage
(Protection of X-ray tube)
The automatic exposure blockage prevents premature
triggering of a new exposure.
After the exposure key has been actuated, the automatic
cool-down pause is indicated.
The LED over key R continues flashing until the cool-off
time has elapsed.
The exposure cannot be triggered until the LED over key
R has gone out.
Example see page 9.
Error Message E ...
Messages such as E2/01 in the kV/mA field indicate
errors.
The LED over the R key blinks.
All unit functions are blocked.
• Press the key R on the Multitimer to reset display
(poss. more than once).
• If the error message is still displayed turn OFF the unit
and then turn it ON again.
If error message has disappeared, all unit functions are
normal again.
For List of Error Messages with description of reme-
dies see page 34