List Of Error Messages; Error Messages - Sirona ORTOPHOS 3 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

11 List of Error Messages

11
List of Error Messages
11.1

Error Messages

Error message E..
One of the keys on the Multitimer was pressed
E1 01
during start-up of the unit.
Faulty communication with the unit.
E1 03
Appears after pressing exposure key.
E2 01
X-ray head overheated.
Cool-down periods ignored.
Faulty communication with the Multitimer.
E2 03
Zero-Power was re-initialized.
E2 04
Max. radiation time of program exceeded.
E2 10
Appears after pressing the exposure key,
E2 20
e.g. if X-ray room door contact not closed.
Exposure lead in Multitimer cable damaged.
Invalid data in data memory.
E2 35
E3 09
Movement of height adjustment obstructed.
Film cassette swings out of panorama exposure
E3 36
position during exposure.
34
Remedy
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display. Freely pro-
grammed values (see page 30) are canceled. Please repro-
gram (see page 28).
Press R key on Multitimer to reset display.
Close X-ray room door.
Press R key on Multitimer to reset display.
If error reappears, poss. break in cable.
Press R key on Multitimer to reset display.
Check that height adjustment can move freely.
Press R key on Multitimer to reset display.
WARNING
If this error message occurs repeatedly during the mo-
tor-driven up and down movement of the rotary unit, es-
pecially when no patient is positioned, switch off the unit
immediately and inform your service engineer without de-
lay.
Swing panoramic cassette holder into panorama position.
Press R key on Multitimer to reset display.
In case of an error message not contained in this list,
switch the unit OFF and ON again.
CAUTION
If an error is displayed again after switching the unit off
and then on again, please contact your service techni-
cian.
Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS 3
60 82 262 D 3285
D 3285.201.04.06.02

Publicidad

Tabla de contenido
loading

Tabla de contenido