Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS 3
6.3
Selecting Data at the Multitimer
I I.3
c
P I
s
66
I0
P I
66
I0
60 82 262 D 3285
D 3285.201.04.06.02
c
s
c
s
Select the Exposure Program
by pressing the – + keys.
The exposure program selected, e.g. P1, and the corre-
sponding maximum exposure time are alternatively
shown on the digital display.
Manual
Select Exposure Data
by tapping one of the four patient symbol keys.
The LED above the key will then light up, and the respec-
tive kV/mA matched pair will appear on the digital dis-
play.
The exposure data can be modified manually with the
– + keys.
The LED over the patient symbol key is then no
manual
longer illuminated.
The mA value is the same for all kV values (see section
"Program Values").
The kV/mA matched values for the patient symbol keys
are factory programmed. Should you need to modify
these values, see the chapter entitled "Programming".
6 Operation
25