Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS 3
9.2
Program Values USA
Index 2A
Program
Program
duration
approx.
P1
21s
P6.1 + P6.2
21s + 21s
P11
18s
Possible kV/mA value pairs – manually selectable with the two smaller patient symbols
60
60
60
62
6
7
10
10
Possible kV/mA value pairs – manually selectable with the two larger patient symbols
60
62
64
66
10
10
10
10
Film and Processing Notes
60 82 262 D 3285
D 3285.201.04.06.02
Exposure
Factory-programmed values
time
with a film density of 03
11.3s
66/10
70/10
7.8s
68/10
72/10
11,3s
66/10
70/10
64
66
68
70
10
10
10
10
68
70
72
74
10
10
10
10
Intensifying Screen Kodak Lanex Regular (sensitive to
green) with film Kodak T-Mat G, Agfa Ortholux. Differ-
ences in film density depending on film and development
tolerances can be compensated by changing the posi-
tion of the density switch (see page 28).
From 03 to 02 when the density is too high, from 03 to 04
when the density is too low.
Freely programmed values or
values with other film density: ....
74/10
78/10
76/10
80/10
74/10
78/10
These values serve only as user reference times.
By actuating one of the two smaller or one of the two
larger patient symbols, you can change between the two
kv/mA pair groups. By actuating the kV/mA + - keys, you
can select the individual values of the kV/mA groups.
72
74
76
10
10
10
76
78
80
10
10
10
The programmed values apply to Kodak T-Mat G and
Agfa Ortholux films in combination with the Kodak Lanex
Regular intensifying screen.
Process according to the manufacturer of the developing
equipment's guidelines or follow instructions on the
chemicals.
9 Program Values
– please enter here –
kV
mA
kV
mA