Assembling the Electric Actuator to the Boom
NOTE: For this procedure, refer to FIGURE 2.5.
1. Remove the shoulder bolt, washer and
nut from the mounting bracket on the
boom assembly.
2. Unpack the pinch guard from the
patient lift carton.
NOTE: The bottom of the electric actuator
assembly will already be assembled to the mast
mounting bracket.
3. Cut the plastic‐wrap that secures the
boom and mast together.
4. Lift‐up on the boom and place it on
your left shoulder.
5. Let the actuator rest on your right‐side of your chest and rotate the shaft extension of
the actuator assembly until it lines‐up with the mounting holes in the boom assembly.
6. Place the pinch guard over the shaft extension of the actuator.
7. Align the holes of the boom assembly mounting bracket with those of the actuator and
insert the bolt. Secure with nut.
NOTE: Be sure that the bolt is completely through the holes of the boom assembly mounting
bracket and the actuator assembly. The boom assembly will pivot easily if the mounting hardware
is aligned properly when the boom assembly is secured to the mast.
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
8. Plug in the electrical connectors from the electric actuator to the bottom of the battery
assembly.
9. Plug the pendant control into the bottom of the battery assembly.
NOTE: The bottom of the actuator assembly will already be assembled to the mast mounting bracket.
Installing the Shifter Handle
NOTE: For this procedure, refer to
FIGURE 2.6.
1. Remove the shifter handle from the
packaging carton.
2. Line‐up the shifter handle threads with
threaded opening in the base.
3. Turn the shifter handle clockwise and
securely tighten into the base.
Part No 1078987
FIGURE 2.5 Assembling the Electric
Actuator to the Boom
CAUTION
Shifter Handle
FIGURE 2.6 Installing the Shifter Handle
19
ASSEMBLY
Mounting Bracket
Pinch Guard
NOTE: Turn 90°
Threads
Opening in Base
Manual/Electric Portable Patient Lift