Replacing The Swivel Bar - Invacare RHL450 Manual Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MAINTENANCE

Replacing the Swivel Bar

WARNING
After the first year of use, the hooks of the swivel bar and mounting brackets of the
boom should be inspected every six months to determine the extent of wear. If
these parts become worn, replacement must be made.
NOTE: For this procedure, refer to FIGURE 7.4.
1. The swivel bar comes attached to the boom with bolt, nut, washers, and pinch guard.
NOTE: Washers should be placed against both sides of the boom mounting bracket to eliminate 
damage.
NOTE: The swivel bar has three hooks on both ends to attach the sling that supports the patient 
during lift.
2. Remove existing hardware and replace the swivel bar.
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
DETAIL "A" - SWIVEL BARS FOR RELIANT LIFTS BEFORE 1/1/06
600 lbs Swivel Bar
450 lbs Swivel Bar
DETAIL "B" - SWIVEL BARS FOR RELIANT LIFTS AFTER 1/1/06
600 lbs Swivel Bar
450 lbs Swivel Bar
FIGURE 7.4 Replacing the Swivel Bar