Opening the Legs
NOTE: For this procedure, refer to FIGURE 3.2.
1. Stand at the rear of the patient lift and grasp the shifter handle with one hand and
place the opposite hand on the steering handle of the mast for balance.
NOTE: The shifter handle MUST lock into its mounting slot to lock the legs in the full open
position.
WARNING
If the shifter handle is not positioned
completely into its mounting slot, DO
NOT use the patient lift until shifter
handle is properly seated and the legs of
the patient lift are locked in place. Oth-
erwise, injury and/or damage may occur.
2. Pull the shifter handle away from the patient lift and then to your right.
Raising/Lowering a Manual/Hydraulic Lift
Invacare does not recommend locking of the rear casters of the patient lift when
lifting an individual. Doing so could cause the lift to tip and endanger the patient and
assistants. Invacare does recommend that the rear casters be left unlocked during
lifting procedures to allow the patient lift to stabilize itself when the patient is ini-
tially lifted from a chair, bed or any stationary object.
There are two controls on the pump assembly:
• The control valve
• The pump handle
Raising the Hydraulic Lift
NOTE: For this procedure, refer to
FIGURE 3.3.
The control valve must be in the closed
position (control valve positioned towards
pump handle) to move the pump handle
up and down to elevate the boom and the
patient until the boom locks in the notch of
the bracket.
Part No 1078987
FIGURE 3.2 Opening the Legs
WARNING
Pump Handle
Closed
Control Valve
FIGURE 3.3 Raising the Hydraulic Lift
21
OPERATION
Open
Manual/Electric Portable Patient Lift
Pump
Handle