MAINTENANCE
Manual/Hydraulic
Mast
Bushing
FIGURE 7.2 Maintaining the Manual/Hydraulic Pump - Replacing a Manual/Hydraulic Pump -
Checking and Tightening Mast Pivot Bolt
NOTE: For this procedure, refer to FIGURE 7.3.
1. Lift up the back of the rubber boot and
slide it off the mast along the boom.
2. Check that the bolt is through the
bracket and the locknut is tight and
secure.
3. If needed, do one or more of the
following:
• tighten locknut and back‐off the
locknut 1/8 of a turn.
• replace the locknut.
4. Reposition the rubber boot.
Manual/Electric Portable Patient Lift
Boom
Pinch
Guard
Bushing
Replacing the Electric Actuator
Bolt
Mast
FIGURE 7.3 Checking and Tightening Mast
38
Electric
Mast
Rubber Boot
Washer
Bracket
Pivot Bolt
Boom
Pinch
Guard
Boom
Spacer
(if equipped)
Locknut
Part No 1078987