Activating A Mechanical Emergency Release; Primary Emergency Release; Performing An Emergency Stop - Invacare RHL450 Manual Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

OPERATION

Activating a Mechanical Emergency Release

Primary Emergency Release

NOTE: For this procedure, refer to FIGURE 3.6.
All lift actuators are equipped with a 
Control
mechanical Emergency release. The 
Box
mechanical release will enable the actuator 
to retract without power. The actuator will 
only retract while under load and the 
mechanical Emergency release is pulled. 
Emergency
The release is colored reddish orange with 
Release
the word Emergency spelled out in white. 
Hole
To activate the primary emergency release, 
insert a pen into the hole labeled 
"emergency" on the control box of the lift 
and push down on the boom at the same 
FIGURE 3.6 Primary Emergency Release
time.
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.7.
NOTE: It is recommended that the primary 
Push DOWN
emergency release be used. The secondary 
on Boom
emergency release is only a back‐up to the 
primary emergency release.
In cases where the primary release is either 
not functioning or unreachable, a 
Pull UP on
secondary emergency release may be used.
EMERGENCY
Ring.
To activate the secondary release, pull up 
on the EMERGENCY ring and push down 
on the boom at the same time.
FIGURE 3.7 Secondary Emergency Release

Performing an Emergency Stop

NOTE: For this procedure, refer to FIGURE 3.8.
Press the RED button on the control box in to stop the lift arms and patient from raising 
or lowering.
Rotate CLOCKWISE to
Disengage Emergency Stop
Control Box
Press IN to
Stop Boom
FIGURE 3.8 Performing an Emergency Stop