OPERATION
Activating a Mechanical Emergency Release
Primary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.6.
All lift actuators are equipped with a
Control
mechanical Emergency release. The
Box
mechanical release will enable the actuator
to retract without power. The actuator will
only retract while under load and the
mechanical Emergency release is pulled.
Emergency
The release is colored reddish orange with
Release
the word Emergency spelled out in white.
Hole
To activate the primary emergency release,
insert a pen into the hole labeled
"emergency" on the control box of the lift
and push down on the boom at the same
FIGURE 3.6 Primary Emergency Release
time.
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.7.
NOTE: It is recommended that the primary
Push DOWN
emergency release be used. The secondary
on Boom
emergency release is only a back‐up to the
primary emergency release.
In cases where the primary release is either
not functioning or unreachable, a
Pull UP on
secondary emergency release may be used.
EMERGENCY
Ring.
To activate the secondary release, pull up
on the EMERGENCY ring and push down
on the boom at the same time.
FIGURE 3.7 Secondary Emergency Release
Performing an Emergency Stop
NOTE: For this procedure, refer to FIGURE 3.8.
Press the RED button on the control box in to stop the lift arms and patient from raising
or lowering.
Rotate CLOCKWISE to
Disengage Emergency Stop
Control Box
Press IN to
Stop Boom
FIGURE 3.8 Performing an Emergency Stop