Low Battery Warning
If battery warning alarm activates while using the device, leave the area immediately as the end of
battery life is approaching. Failure to follow this warning can result in serious personal injury or death.
Fig. 4
Battery Warning
The duration of remaining device operation during a Low Battery Warning depends on ambient
temperatures, battery condition alarm status. Nominal battery life is 30-60 minutes after the
Battery Warning activates.
When the device goes into battery warning the:
-
battery life indicator continuously blinks
-
alarm sounds and alarm LEDs flash every 30 seconds
-
Safe LED no longer flashes
-
device continues to operate until it is turned OFF or battery shutdown occurs.