Remove the battery pack.
(3)
Remove the two remaining case screws, and remove the case front.
(4)
Gently remove the sensor to be replaced.
(5)
Carefully align the new sensor contact pins with the sockets on the printed circuit board.
(6)
Press the new sensor into place.
(7)
Note the position restrictions in the table above.
Adapter (part no. 10110183) is required for XCell usage in position 3.
If a sensor is removed and will not be replaced, be sure to install a sensor plug in its place
in order to maintain correct device function.
The plug for XCell positions is P/N 10105650.
The Series 20 plug is P/N 10088192.
(8)
Reinstall the sensor gasket in the case front.
(9)
Attach front case and tighten two case screws using 5.5 in-lbs of torque.
(10)
Attach the battery pack and tighten the two battery pack screws using 5.5 in-lbs of torque.
If a change in XCell Sensor configuration is detected during the device turn-ON process:
-
The "ACCEPT?" prompt appears on the display
-
The  button accepts the sensor configuration
-
The  button rejects the sensor configuration; the device is not operational.
When an XCell sensor is replaced, the device automatically enables the sensor after the change has
been accepted. If a Series 20 sensor is replaced, it must be manually enabled (→ chapter 3.5
SETTING SENSOR OPTIONS).
If the oxygen sensor was replaced, see chapter 3.2 regarding the oxygen reading display.
(11)
Allow sensors to stabilize at least 30 minutes before calibration.
(12)
Calibrate device before use.
Calibration is required after a sensor is installed; otherwise, the device will not perform as expected
and persons relying on this product for their safety could sustain serious personal injury or death.
WARNING
ALTAIR 5X/ALTAIR 5X IR
Maintenance