MSA ALTAIR 5X Manual De Uso página 62

Detector de gases múltiples
Ocultar thumbs Ver también para ALTAIR 5X:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ALTAIR 5X IR
Manufacturer:
Product:
IECEx-Type
Examination Certificate:
Type of protection:
Performance
Marking:
Li-Ion
Special Conditions:
The model ALTAIR 5X shall be charged by Manufacturer's chargers only (0 - 45 °C) and openedin a
non-hazardous area.
When using the ALTAIR 5X or ALTAIR 5X IR in a hazardous area, the device should be worn or carried
on the body. Do not store the device in a hazardous location. This prevents the possibility of the device
building up an electrostatic charge.
The antenna used for activation of the internal RFID tag with the RF radiation power shall not exceed
6 W for Group I and 2 W for Group IIC.
In the event of a combustible sensor overrange, the device should be exposed to fresh air for a
minimum of 20 minutes. Following this, a Zero Calibration procedure should be performed.
The warm-up time for oxygen is up to 180 sec.
The alarm set points are not applied for measuring oxygen inertisation and it shall be taken into
account.
Capacitance:
5X Alkaline Battery Pack Screws:
5X Rechargeable Battery Pack D-Ring:
5X Charge contact pins:
5XiR Rechargeable Battery Pack D-Ring:
5XiR Charge contact pins:
Mine Safety Appliances Company, LLC
1000 Cranberry Woods Drive
Cranberry Township, PA 16066 USA
ALTAIR 5X IR
IECEx TSA 09.0014X
IEC 60079-0:2004, IEC 60079-1:2003
IEC 60079-11:2006
none
Ex d e ia I Ma
EX d e ia IIC T4 Gb
Ta = -40 °C to +50 °C
Um ≤ 6,7 V DC
IP 65
6 pF
26 pF
16 pF
33 pF
24 pF
ALTAIR 5X/ALTAIR 5X IR
Certification
US