Operation

Disinfect or autoclave contaminated surfaces of the
curing light, light probe and anti-glare cone before each
use. Furthermore, make sure that the light output permits
adequate polymerization. For this purpose, check the
light probe for dirt or damage and use the integrated
light meter to test the light output.
Switching the handpiece on and off (stand-by mode)
If the handpiece is not used for approximately 4 minutes,
it changes to the energy-saving stand-by mode and auto-
matically switches off. Renewed pressing of the start but-
ton results in the stand-by mode being switched off and
the handpiece switched on again. At the same time, the
display is activated.

Selecting the curing program

The desired curing program is selected with the program selection buttons. The display changes accordingly
(see Indicators on the handpiece).
bluephase C8 is equipped with the following 3 curing programs to match the requirements of different indications:
Display
Curing Program
Intensity curve
HIP (High Power)
800 mW/cm
0
LOP (Low Power)
650 mW/cm
0
SOF (Soft start)
800 mW/cm
650 mW/cm
0
5
1) The stipulations apply to the 13>8 mm Power-Booster light probe supplied with the kit
8
Application
Consistently high light intensity
for the polymerization of
2
restorative and cementation
materials in direct and indirect
restorations.
t [s]
Reduced light intensity with
reduced heat build-up for the
polymerization of adhesives, liners
2
and restorative materials in areas
near the pulp when restoring
Class V cavities.
t [s]
Step-by-step increase of the light
intensity with reduced shrinkage
2
stress and reduced temperature
build-up for the polymerization of
2
restorative materials.
t [s]

Settings

The curing program and curing time can be individually
set. On initial start-up, the device will operate with the
following presets :
– HIP curing program (High Power)
– Curing time: 20 seconds
For the other curing programs, the following curing times
have been preset:
– LOP (Low Power): 10 seconds
– SOF (Soft start): 30 seconds
Stipulated curing time
of selected materials
10 seconds:
– Adamant
– Indirect restorations with dual cure cement
– Per segment: occlusal, mesial, distal
– Indirect restorations with the light cure cements
Variolink II and Variolink Veneer
– Per segment: occlusal, mesial, distal
– Per millimeter thickness of restoration
i.e) 2 mm thick restoration, base only
(10 seconds x 2) = 20 seconds per segment
20 seconds:
– Generelly for composites such as
Tetric EvoCeram, Tetric EvoFlow, Tetric N-Ceram, Heliomolar
40 seconds:
– Compoglass F | Flow
10 seconds:
– AdheSE | One
– Excite | DSC
20 seconds:
– Syntac
– Heliobond
30 seconds:
– Generelly for composites such as
Tetric EvoCeram, Tetric EvoFlow, Tetric N-Ceram, Heliomolar
1)