Connecting the Visum
®
to SIDNE
SIDNE can control up to four Visum LED Surgical Lights. To do this, a second
power control box is necessary. Use the Expansion port on the first power supply
box to control the additional two lights.
To enable SIDNE to control four surgical lights, do the following:
1.
Power on the lights and camera.
2.
Power on SIDNE. Wait for SIDNE to say, "SIDNE ready. "
3.
Connect the USB A – to – B (male) cable from an available SIDNE
device port to the SIDNE port on the first power supply box.
4.
Insert a second Stryker USB A – to – B cable into the expansion port.
5.
Connect the 50-foot cable (leading from the expansion port) to the
SIDNE® port on the second power supply box (see diagram).
6.
SIDNE should say, "O.R. Lights connected" within 20 seconds after the
Visum LED Light and Camera and SIDNE have been connected. If this
does not occur, refer to the Troubleshooting section of this manual for
instructions.
12
LED Light and Camera
™