6
Prolystica 2X Neutral
Detergent at 1/8 oz/
gal. at 35-40°C is
recommended for
cleaning.
7
Cleaning was
validated using
Neodisher Mediclean
Forte
at 5 ml/L at
TM
30°C.
5. Soak in Detergent
Note: In the U.S., use a non- enzymatic detergent. In
the EU, use Neodisher Mediclean Forte.
U.S. only instructions
• Prepare a non-enzymatic detergent solution
lukewarm tap water according to manufacturer's
recommendations.
• Immerse the handpiece in the detergent solution,
ensuring the solution contacts all inner and outer
surfaces of the handpiece. Use a syringe to inject
lumens with 50mL of solution.
• Soak the handpiece in the solution for 15 minutes.
EU only instructions
• Prepare detergent solution
according to manufacturer's recommendations.
• Immerse the handpiece in the detergent solution,
ensuring the solution contacts all inner and outer
surfaces of the handpiece. Use a syringe to inject
lumens with 50mL of solution.
• Soak the handpiece in the solution for 1 minute.
in lukewarm tap water
7
in
6
EN-21