Automated
Cleaning
12
Cleaning was
validated using
Enzol® at 1 oz/gal. at
35-40°C.
13
Cleaning was
validated using
Neodisher Mediclean
Forte
at 5 ml/L at
TM
30°C.
9. Inspect
Visually inspect the handpiece, including all
internal surfaces, for remaining soil. Use an
endoscopic camera and endoscope if necessary
to see the inner surface of the lumen.
If soil remains, repeat the manual cleaning
procedure, focusing on those areas.
1. Wipe
Wipe excess soil from the handpiece with disposable
paper towels.
2. Soak in Detergent
Note: In the U.S., use an enzymatic detergent. In the
EU, use Neodisher Mediclean Forte.
U.S. only: Prepare an enzymatic detergent
solution
manufacturer's recommendations.
EU only: Prepare detergent solution
lukewarm tap water, according to manufacturer's
recommendations.
Wipe the entire surface of the handpiece using a
soft clean cloth dipped in the detergent solution.
Immerse the handpiece in the detergent solution,
ensuring the solution contacts all inner and outer
surfaces of the handpiece. Use a syringe to inject
lumens with 50mL of solution.
Soak the handpiece in the solution for 15 minutes.
with lukewarm tap water, according to
12
with
13
EN-23