Disinfection
21
CIDEX® Activated
was validated for
disinfection efficacy.
EN-26
Warning
1.
Disinfect the device in a disinfecting solution that
has ≥ 2.4% glutaraldehyde as an active ingredient.
Prepare the solution according to the manufacturer's
instructions.
2.
Immerse the device in the disinfecting solution, filling
all lumens, for 45 minutes at 25°C.
3.
Rinse the device in a bath of sterile water. Immerse the
device, shake it under water, and allow it to soak for
one minute. Repeat the rinse two more times, using a
fresh bath each time.
4.
Dry all parts with sterile, lint-free cloth immediately
after rinsing.
Detailed Thermal Disinfection
Instructions:
1.
Disassemble the instruments according to the
"Cleaning" section included in this manual.
2.
Place the disassembled instruments into the washer
disinfector.
3.
Program the washer disinfector using the following
parameters:
Thermal Disinfection Parameters
Recirculation Time
1 minute
5 minutes
Not to be used in place of
sterilization.
21
Water Temperature
93°C
90°C