17
Cleaning was
validated using
Enzol® at 1 oz/gal. at
35-40°C.
18
Prolystica 2X
Neutral Detergent at
1/8 oz/gal. at 35-40°C
is recommended for
cleaning.
Pre-Wash
Enzyme Wash
Wash 1
Rinse 1
Dry
19
Cleaning
was validated
using Neodisher
Mediclean Forte
TM
at 5 ml/L at 30°C.
Pre-Wash
Wash 1
Rinse 1
Dry
20
Drying was
validated with
pressurized air at
20 psi.
5. Automated Wash
Place the handpiece in an automated washer, on
an incline to aid in drainage.
Program the washer with the following
parameters, then activate the wash:
Recirculation
Time
2 Minutes
2 Minutes
2 Minutes
2 Minutes
7 Minutes
Recirculation
Time
2 Minutes
2 Minutes
2 Minutes
7 Minutes
6. Dry
Dry the handpiece using a clean cloth. Filtered
pressurized air can be used to assist drying.
7. Inspect
Visually inspect the handpiece, including all
internal surfaces, for remaining soil. Use an
endoscopic camera and endoscope if necessary
to see the inner surface of the lumen.
If soil remains, repeat the automated cleaning
procedure, focusing on those areas.
U.S. Parameters
Water
Temperature
Cold
Hot
60°C (set point)
Hot
115°C
EU Parameters
Water
Temperature
Cold
60°C (set point)
Hot
115°C
Detergent
Type
NA
Enzymatic
17
Non-Enzymatic
18
NA
NA
Detergent
Type
NA
Neodisher
Mediclean Forte
NA
NA
20
EN-25
19