Function and Limitations of Use

The AutoRep
pensing of liquids. Volumes from 2 µL to 5 mL can be dispensed with the
highest precision and accuracy through combinations of the syringe size
used and adjustment of the stroke setting on the stroke-setting wheel.
Limitations of Use
This instrument is designed for dispensing samples, observing the following
limits:
– 59°F to 104°F (+15°C to + 40°C)
(of instrument and reagents – other temperatures on request)
– vapor pressure up to 500 mbar
– viscosity:

Operating Exclusions

During proper operation, the dispensed sample only comes in contact with
the syringe and not with the AutoRep
The user has to ensure the compatibility of the instrument with the intended
application. In case of third-party dispenser syringes, the user must check
their compatibility and perform a function and volume test before use.
The instrument should not be used for dispensing liquids that attack
polypropylene (e.g., syringe cylinder), polyethylene (syringe piston),
LCP (syringe piston 0.1 mL), or PC/PBT and polycarbonate.
Note:
Instrument and syringe are not autoclavable (sterile Encode syringe, see
page 17). Syringes are disposable. For decontamination of the instrument
you may use commercially available solutions as long as they do not attack
PC/PBT and PC.
6
S is a repetitive pipette for quick and simple repetitive dis-
20 mPa s with 50 mL syringes
260 mPa s with 5 mL syringes
977 mPa s with 1 mL syringes
S.