The possible dispensing volumes of the AutoRep
These are found beneath the transparent clip on the back of the instrument.
There are 9 stroke setting positions, from 1 to 5.
1.
Look for the desired dispensing volume (Volume) in the table.
2.
Use the stroke setting wheel to set to the proper stroke number (Setting)
for the volume that corresponds to the syringe size.
3.
Dispensing steps: depending on the syringe capacity (mL), the same
desired dispensing volume per step can be delivered a different number
of times (see example below). The smaller the number of steps, the
higher the accuracy of the dispensed volume.
Example:
Important!
Desired dispensing volume per step: 200 µL
Syringe sizes to use:
2.5 mL = stroke setting 4 = max. 11 dispensing steps
5 mL = stroke setting 2 = max. 24 dispensing steps
10 mL = stroke setting 1 = max. 49 dispensing steps
The volumes listed in the table are the only options avail-
able. Since the first dispensing step must be discarded,
the number given in the table is always 1 step less.

Setting the Volume

S are listed in tables.
9