To Clean SPIDER2™
1. Remove all single use components and accessories. Discard single use components.
2. Clean and process each accessory.
Note: Refer to cleaning and sterilization instructions for accessories: 79-99-04 Cleaning
and Sterilization Instructions.
3. Wipe entire SPIDER2 with mild detergent / disinfectant, taking care to avoid fluid entry into
Switch Drape or Foot Pedal Receptacles.
4. Wipe battery with mild detergent / disinfectant.
Note: Avoid the use of excessive amounts of liquid during cleaning.
5. Dry SPIDER2 with a cloth or allow to air dry.
6. DO NOT immerse any portion of SPIDER2 in liquid.
Note: If moisture enters the equipment it can cause corrosion, and damage electrical
and/or mechanical components.
7. DO NOT autoclave SPIDER2.
8. DO NOT use ethylene oxide to sterilize SPIDER2.
Storage
1. Turn OFF and DEPRESSURIZE the SPIDER2 by disconnecting the Foot Pedal and/or
Switch Drape and pressing the Power Switch to the DEPRESSURIZE Position for 1 second
(Figure 4).
CAUTION: Do not store SPIDER2 in locked position as this will drastically reduce the
life expectancy of device. Depressurize before storing by turning Power Switch to the
DEPRESSURIZE Position for 1 second and then allowing the Switch to return to the OFF Position.
2. Store the SPIDER2 on the SPIDER2 Cart by hanging the Hinge Joint over the support hook.
Maintenance
Only trained individuals should perform maintenance on SPIDER2. Contact Smith & Nephew
to determine which maintenance can be performed on site.
8
126-99-01 Rev. H
SPIDER2 Specifications*
* Specifications are approximate and may vary from unit to unit or as a result of power supply
fluctuations.
Model
REF 72203299 SPIDER2
Length: 1495 mm Width: 180 mm
Size (fully extended)
Height: 200 mm
Weight
30 lbs [13.6 kg] (with battery installed)
IEC 60601-1:1988+A1:1991+A2:1995
Compliance
IEC 60601-1:2005 +A1:2012
IEC 60601-1-2:2015
Type B Applied Part
Equipment type
Power Supply
Internally Powered 24V
Battery Type
NiCd 24V 1.8Ah
REF 72203301 SPIDER2 Battery Pack
Battery Model
Enclosure Protection
IPX1 Drip-proof equipment
Environmental Conditions
Operation
Temperature
10° C–40° C
Relative Humidity
30%–75%
Atmospheric Pressure
70 kPa–106 kPa
EU: Not for general waste
Contact Smith & Nephew for appropriate methods of
Disposal
disposal.
SPIDER2 Battery Charger Specification
IEC 60601-1:1988+A1:1991+A2:1995
Compliance
IEC 60601-1:2005 +A1:2012
IEC 60601-1-2:2015
Power supply
100–240 VAC, 50–60 Hz, Max 0.9 A
SPIDER2™ Limb Positioner
20A
Storage & Transportation
-40° C–70° C
10%–100%
50 kPa–106 kPa