Instructions for Use
Error Messages
Message
Fault: short
circuit
System override
Overvoltage
204/262
Description
After the starting phase the output
connection is shorted (output
current higher than 30 A).
The pump is stopped by an
emergency stop signal provided
via a remote contact.
Controller power supply circuitry is
faulty, or the Controller received a
spike.
Turbo-V 1000HT PCB Controller User Manual / 87-900-906-01
Repair action
Check connections and
shortages between pump and
controller. Press the START
push-button twice to start
the pump.
Remove the controller power
cable and check the
emergency condition. Then
reconnect the power cable
and press the START push-
button twice to start the
pump.
Press the START push-button
twice to start the pump.
Should the message still be
present, call the Agilent
service