*The values for power and voltage are given as numbers, scaled from
0 to 255, where 0 corresponds to 0 V and 255 to the full scale voltage
(100 V) or power (500 W).
When the Request is "G", the Answer will contain a string of 12
characters with the following parameters:
Tab. 8
Bytes
1
2-3
4
5-8
9
10
11
12
When the configuration parameters have to be changed, send a
Request string with "H" characters of the following type:
"H" + .... DATA .... + CRC
The CRC is calculated on the string consisting of
H + DATA.
The following parameters can be changed:
Turbo-V 1000HT PCB Controller User Manual / 87-900-906-01 (F)
Meaning
Voltage*
Frequency
Cycle #
R1 Status
R2 Status
CRC
Meaning
Speed adjust value (in KRPM)
Pump cycles number (integer coded in 2 bytes)
Speed threshold
Run up time in seconds (long coded in 4 bytes)
Deat time (0 = NO 1 = YES)
Soft Start mode (0 = NO 1=YES)
Water cooling (0 = NO 1=YES)
CRC
Technical Information