17
Technical Information
Optional Serial Port
Request is an ASCII character identifying the action that must be
performed by the controller or the requested information.
Allowed ASCII characters:
The command "N", "O", "P", "Q", "R" and "S" are always acknowledged
in any chosen operation mode.
CRC corresponds to the sum (with inverted sign) of all the preceding
bytes.
e.g., the START command "A" in ASCII code = 41;inverted it will be:
FF + 1 - 41 = BF.
220/262
"A" = START
"B" = STOP
"C" = Low Speed ON
"D" = Low Speed OFF
"E" = Request for operational parameters
"F" = Pump times zeroing
"G" = Parameters reading
"H" = Parameters writing
"I" = Request for operating status
"J" = Request for numerical reading
"K" = Request for counters reading.
"L" = Software version
"N" = Print to host serial mode change
"O" = Host to print serial mode change
"P" = Front panel operation mode selection
"Q" = Remote operation mode selection
"R" = RS 232 operation mode selection
"S" + "number" = Baud rate selection
Turbo-V 1000HT PCB Controller User Manual / 87-900-906-01
(F)

Publicidad

loading

Este manual también es adecuado para:

969-9879