3.1 General Maintenance Information

It is recommended that periodic inspection and performance testing be completed by a qualified facility
biomedical technician to ensure continued safe and effective operation.

3.2 Cleaning

• Unplug the Bovie
Smoke Shark Surgical Smoke Evacuator.
®
• Wipe Bovie
Smoke Shark Surgical Smoke Evacuator with a damp cloth containing mild disinfectant
®
solution or soapy water.
• Wipe Bovie
Smoke Shark Surgical Smoke Evacuator with a dry clean cloth.
®
• Do not steam sterilize.

3.3 Periodic Inspection

The Bovie
Smoke Shark Surgical Smoke Evacuator should be visually inspected at least once per year.
®
The inspection should include checks for:
• Damage to the Power Cord or Power Inlet Module.
• Obvious external or internal damage to the Bovie
FUSES
Two 4 AMP fuses for 100/120 Bovie
220/240 Bovie
Smoke Shark Surgical Smoke Evacuators are located within the housing of the system.
®
The fuses electrically protect both the Bovie
from damage or injury. If the Bovie
is an electrical surge, fuses will break and the Bovie
operate. No parts are user serviceable. When the Service LED Light Indicator illuminates, please contact
Customer/Technical Services.

3.4 Troubleshooting

PROBLEM
1. Bovie
Smoke
®
Shark Surgical
Smoke Evacuator
is on but suction is
minimal or none.
2. Bovie
Smoke
®
Shark Surgical
Smoke Evacuator
does not function
even though
Suction On/
Standby Button is
depressed.
Customer/Technical Services: 800.251.3000 (United States) or +1 615.964.5532 (International)
Smoke Shark Surgical Smoke Evacuators or two 2 AMP fuses for
®
Smoke Shark Surgical Smoke Evacuator and the operator
®
Smoke Shark Surgical Smoke Evacuator is overheated or if there
®
POTENTIAL CAUSE
1. Filter is not seated
completely.
2. Filter is clogged.
3. Vacuum tubing is
clogged.
4. Motor is obstructed.
1. Not plugged into an
electrical outlet.
2. Fuses are blown.
3. Electronic Bovie
Smoke
®
Shark Surgical Smoke
Evacuator failure.
4. Filter Life has expired or
invalid filter installed.
MAINTENANCE: SECTION 3.0
Smoke Shark Surgical Smoke Evacuator.
®
Smoke Shark Surgical Smoke Evacuator will not
®
CORRECTIVE ACTION
1. Re-install the filter, press firmly into place.
2. Replace the filter with the manufacturer's filter. (SF06)
3. Replace vacuum tubing with manufacturer's products.
4. Call Symmetry Surgical Customer Service.
1. Check the grounded electrical outlet connection and
the power receptacle located at the in the bottom of
the Bovie
Smoke Shark Surgical Smoke Evacuator.
®
2. Call Symmetry Surgical Customer Service.
3. Call Symmetry Surgical Customer Service.
4. Replace the filter.
P000027200 Rev A Bovie
Page 17
Smoke Shark III Operator's Manual
®